5866RB1A1G4M2NA is a configured Brooks Instrument 5866RT-family pressure controller/flowmeter identifier observed in stored auction records. The 5866RT platform controls upstream or downstream pressure from a remote pressure transducer signal while also measuring mass flow through the control valve. Exact gas, full-scale flow, seal package, valve behavior, pressure-transducer range, fitting package, and calibration details must be verified from the complete model code and physical label
Configurations vary. Verify the exact unit before choosing a replacement.
5866RT family supports elastomer-sealed and metal-sealed configurations
Function
Pressure-control function
Maintains a desired upstream or downstream pressure by comparing a remote pressure transducer signal to the setpoint and modulating the integral control valve
Flow-measurement function
Measures mass flow through the control valve and provides a 0-5 V signal proportional to mass flow
Mass-flow-controller mode
5866RT can be configured as a mass flow controller for calibration or test purposes
Pressure
Control architecture
Remote pressure sensor input with adjustable integral PID control electronics and top-access tuning switches
Pressure-control range
100:1 when used with a remote transducer that has a 0-10 V output
Pressure response time
Less than 0.8 seconds typical for a 0-100% command step with less than 2% pressure overshoot; actual response depends on system design
Remote pressure sensor compatibility
Suitable for pressure sensors with 0-5 V or 0-10 V output signals
Flow
Flow range
Any range from 0-3 sccm to 0-30,000 sccm nitrogen equivalent at standard SEMI reference conditions
Flow output signal
0-5 Vdc, 1 kOhm maximum load
Flow accuracy
+/-1.0% full scale including linearity at calibration conditions for 3-10,000 sccm nitrogen equivalent; +/-1.5% full scale for flow rates greater than 20 slpm
Flow repeatability
+/-0.25% of rate
Flow temperature coefficient
0.1% full scale per degree C
Controls
Pressure setpoint signal
0-5 Vdc or 0-10 Vdc command, with 5 Vdc reference output available for setpoint generation
Input/output offset
0.2% full scale
Valve override
Electrically activated valve override is part of the 5866RT control platform
Electrical
Electrical connector
15-pin D-connector, DA-15P
Power requirement, normally closed valve
3.5 W; +15 Vdc +/-5% at 35 mA and -15 Vdc +/-5% at 180 mA
Power requirement, normally open valve
10.5 W; +/-15 Vdc +/-5% at 350 mA
Ratings
Operating temperature range
32 to 150 degrees F (0 to 65 degrees C)
Pressure Equipment Directive context
Brooks data sheet lists PED 97/23/EC Sound Engineering Practice
Materials
Wetted parts
Standard wetted parts are 316L and 316L VAR stainless steel
Valve seat options
Standard Viton fluoroelastomer or metal valve seat depending on configured model code
Seal options
External and internal seals can be fluoroelastomer or metal, with Buna-N, Teflon, or Kalrez options depending on configuration
Leak integrity
Elastomer seal: 1 x 10^-9 atm cc/sec helium maximum; metal seal: 1 x 10^-10 atm scc/sec helium maximum
Connections
Standard mechanical connections
1/8 in or 1/4 in stainless steel compression fittings on standard 5866RT configurations
Optional mechanical connections
1/4 in VCO or VCR options; 5866RT(M) metal-sealed configuration uses 1/4 in VCR
Installation
Mounting orientation
May be installed in any position
Filter recommendation
Brooks recommends an upstream in-line filter, such as Model 5848 for elastomer 5866RT installations, to help prevent foreign material from entering the meter
Applications
Application domain
Upstream or downstream gas pressure control with simultaneous flow measurement, including vacuum-chamber pressure regulation, inert-gas feed control, process vessel pressure control, and flow-related calibration or test service