5861IA112H2DEA is a configured Brooks Instrument 5861i mass flowmeter identifier. Official Brooks documentation identifies the Model 5861i as a mass flowmeter for accurately measuring gas flow, with standard nitrogen-equivalent ranges from 10 slpm to 100 slpm and 15 to 28 Vdc power. Stored Sonisys auction records and surplus/calibration pages associate this exact identifier with 100 SLM and 15-28 Vdc context; those observed details should be confirmed from the unit label and calibration records before reuse
Unit-specific configuration
Observed configuration
Historical listing records identify this part as 100 SLM. Exact gas, range, fittings, valve configuration, and communication interface must be verified from the physical unit label.
Configurations vary. Verify the exact unit before choosing a replacement.
Application classIndustrial automation and process equipment
CategoryFlow Control
Specifications
Identification
Configured part identifier
5861IA112H2DEA
Model family
Brooks Instrument Model 5861i mass flowmeter
Device type
Gas mass flowmeter with flow sensor and integral electronic signal conditioner
Function
Primary function
Measures gas mass flow and provides a proportional flow output signal
Flow
Official model range context
Brooks 5861i manual lists standard ranges from 10 slpm to 100 slpm nitrogen equivalent, with up to 200 slpm hydrogen
Performance
Accuracy context
Brooks 5861i manual lists +/-1% full scale including linearity at calibration conditions
Repeatability context
Brooks 5861i manual lists repeatability as 0.25% of rate
Response time context
Brooks 5861i manual lists less than 6 seconds response to within 2% of full-scale final value for a 0 to 100% flow step
Pressure
Working pressure context
Brooks 5861i manual lists 1500 psi / 10.34 MPa maximum working pressure
Electrical
Power context
Brooks 5861i manual lists +15 to +28 Vdc at 90 mA, 2.5 watts maximum at 28 V
Output signal context
Brooks 5861i manual lists 0-5 Vdc output plus jumper-selectable 4-20 mA or 0-20 mA current output
Mechanical
Electrical connection
Brooks 5861i manual identifies a DA-15P 15-pin male D-type connector
Gas connection context
Brooks 5861i manual lists standard 3/8 inch compression fittings, with VCO and VCR options available on request
Environmental
Operating temperature context
Brooks 5861i manual lists operating ambient temperature from 5 C to 65 C / 40 F to 150 F
Configuration
Configured-unit rule
Gas, full-scale calibration, fittings, output mode, calibration conditions, and any customer-specific build details vary by configured code and physical unit label
Observed configuration
Stored listing context
Stored Sonisys auction records and surplus pages for this exact identifier use 100 SLM and 15-28 Vdc wording; this is observed listing context, not an official configured-code build sheet