Brooks Instrument Sho-Rate 50 1350E Variable Area Flowmeter
About this part
1350EBB9CJV5A is a configured Brooks Instrument Sho-Rate 50 / Model 1350E glass tube variable area flowmeter. The stored auction record consistently pairs the code with 8900D3J2B1AC and a 0-205 SCFH scale, so Sonisys treats that flow range as observed configured-unit evidence while using the official Brooks 1350E/1355E documentation for family-level construction, performance, installation, and service context
Configurations vary. Verify the exact unit before choosing a replacement.
Application classIndustrial automation and process equipment
CategoryFlow Control
Specifications
Identification
Device type
Glass tube variable area flowmeter
Series family
Sho-Rate 50 Model 1350E / Sho-Rate 150 Model 1355E family
Configured model code
1350EBB9CJV5A
Observed associated code
8900D3J2B1AC appears with this part number in stored auction titles
Flow
Measuring principle
Variable area rotameter using a tapered borosilicate glass metering tube and float
Observed configured scale
0-205 SCFH appears in every stored title for this part record
Performance
1350 Series accuracy
Standard accuracy is +/-10% of full scale
1355 Series accuracy
Standard accuracy is +/-5% of full scale
Repeatability
0.5% full scale
Pressure / Temperature
Standard pressure and temperature rating
200 psig from 33°F to 250°F / 1°C to 121°C
CRN pressure rating
170 psig from 33°F to 250°F / 1°C to 121°C when supplied under CRN certification
Construction
Metering tube material
Borosilicate glass
Float material options
Glass, 316 stainless steel, sapphire, Carboloy, and tantalum
End fitting material options
Chrome plated brass, black anodized aluminum, and 316 stainless steel
Window materials
Clear polycarbonate front window with milk white polycarbonate back window
Connections
Standard connection
Horizontal female 1/8 inch NPT threaded adapters with locknuts for front panel mounting
Optional connections
1/8 inch and 1/4 inch compression fittings, 1/4 inch female NPT, and 1/4 inch ID serrated hose connections
Controls
Integral control options
Standard integral flow control valve, NRS integral flow control valve, and 8800/8900 integral flow controller configurations are supported by the family documentation
Installation
Orientation
Install within 6 degrees of true vertical with inlet connection at the bottom end fitting
Configured-unit rule
Tube, float, scale, connection, seal, valve/controller location, material selections, and exact fluid calibration vary by configured code and unit label
Applications
Application domain
Low-flow gas and liquid indication, purge flow service, panel-mounted rotameter applications, and configured flow-control assemblies